Log entry time 14:39:49 on November 01, 2009
Entry number 297096
This entry is a followup to: 296462
keyword=Compton cavity locked and optimised
The Compton cavity was locked for the first time since the second power
failure. The mirror positions were lost. The following procedure was used.
1) Maximise the reflected photodiod signal (on the scope.)
2) Use the mirrors to move the beam on the QPDs nearer to the last
recorded values in the following order.
a) M1V -> Y2
b) M1H -> X2
c) M2V -> X1
d) M2H -> Y1
3) Once flashing is observed, move the mirrors to maximise the depth of
the reflection dip (on the scop) until locking occurs.
4) Optomise alignment for lock.
The final position was a power of 965 W at QPD positions
Y1=121
X1=511
Y2=18
X2=-604