    User name D.Wang/X.Deng

    Log entry time 02:52:43 on November 3,2009

    Entry number 297225

    This entry is a followup to: 297206

    Followups:

    keyword=RHRS s1 3rd and 4th ADC

    It looks like the reason why we don't see 3rd and 4th ADC for s1 is that the 
    database is wrong. I was taking some test runs and working in the directory 
    happexsp/onlana. In the database for s1 (I'm asumming it's in the sub directory 
    DB/20090602), the ADC channels for the 3rd and 4th PMTs are set to 61 and 62, 
    which is not true. We checked the cables and found that the two PMTs are 
    connected to the 3rd and 4th channel of the ADC(slot 25). So I changed the 
    database (saved the old one as db_R.s1_Nov3_2009.dat) and reanalyzed the data. 
    
    The first graph is for the RHRS run 4735. The 3rd and 4th of the left pads have 
    signal now. However, the shape is different from the others. In order to 
    confirm that we are seeing the correct signal, I changed around the HVs(they 
    were -1900V before change) for these two PMTs and took several runs: 
    run 4736, HV for 3rd PMT is 1800, HV for 4th PMT is 1800
    run 4737, HV for 3rd PMT is 1750, HV for 4th PMT is 1750
    run 4738, HV for 3rd PMT is 1700, HV for 4th PMT is 1720
    And we do see the ADC spectrums shrink as we decrease the HVs, which implies we 
    are really seeing the 3rd and 4th PMTs now, meaning the database is correct now.
    
    The second graph is for the run 4738. The spectrum for the 3rd PMT looks 
    different, but the two PMTs are basically working. 
    
    There's some problem with the left 5th PMT now. Not sure what is the problem. 
    Will check it later. 
    