keyword=R-HRS DAQ warnings From Roc 2, slot 22
Last night apparently something changed in ROC 2, slot 22. Jin found that
after run 4741, the right HRS replay would no longer work and produced
repeated warning messages:
DAQ warning: too many data on crate 2 slot 22 : 64
I looked into the db_cratemap.dat file and found that this was labeled as
an ADC. I remembered looking at ROC 2 and recalling that there was a TDC
in this slot. I changed the db_cratemap file to reflect this. Now the
replay for the right HRS is working.
Someone should double check what module is in this crate and slot.
Here is the line I modified in the cratemap file:
# slot model clear header mask nchan ndata
22 1877 1 0x0 0x0 96 672
A copy of this log entry has been emailed to: rsubedi,jinhuang,pkpan,dwang,deng