• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name D.Wang

    Log entry time 02:00:44 on November 07, 2009

    Entry number 297975

    This entry is a followup to: 297956

    keyword=LHRS DAQ problem persists

    We asked for an escorted access trying to fix the LHRS DAQ problem. I checked the gate signal from the TM module on the scope and didn't see any signal. Then I found there was nothing plugged in the L1A of the TM module. There was an ECL cable labled with "1 DAQ L1A" lying on the floor, both ends unplugged. I tired taking an L1A copy from the delay module and plugging it into the TM module, but it didn't fix the problem, maybe because it's not well timed. Then I gave up due to limited time. We need to fix it during next access.

    A copy of this log entry has been emailed to: rom, rsubedi, xiaochao