User name deng/dwang
Log entry time 07:27:25 on November 7,2009
Entry number 298044
This entry is a followup to: 297771
keyword=boiling test, fan speed part
Since MCC has trouble sending us beam higher than 85uA, we did fan speed at 40,
50, 60Hz at beam currents 10, 20 and 85uA with raster size 6x6 in MCC unit.
Before we did fan speed test, three runs were taken at 85uA with fan speed at
60Hz and raster size 2x2, 4x4 and 6x6 (all mcc unit) to repeat raster size test
done yesterday owl shift. This is because when we did raster size test
yesterday, the right arm Q1 was down while the one in left arm was fine. Rupesh
mentioned that the lumi width would be quite different in this situation
compared to both arm Q1 on or off. This is verified by the test.
1) raster size test results @85uA
run target raster size good pairs AsymNBlumiSum width (ppm)
13774 LD2 6x6 9.8K 557.9
13775 LD2 2x2 10.8K 6635
13776 LD2 4x4 5.8K 1650
while yesterday,
13752 LD2 6x6 6.2K 592.3
13753 LD2 4x4 8.7K 1874
13754 LD2 2x2 13.7K 7361
2) fan speed test, raster size 6x6 in MCC unit
run || target || current || fan speed || #pairs || AsymNBlumiSum width
13774 LD2 85uA 60Hz 9.8K 557.9
13777 LD2 85uA 50Hz 6.7K 840
13778 LD2 85uA 40Hz 7.8K 2088
13780 LD2 20uA 40Hz 6.7K 293.3
13781 LD2 20uA 50Hz 6.6K 280.6
13782 LD2 20uA 60Hz 7.1K 274
13784 LD2 10uA 60Hz 6.9K 448.9
13787 LD2 10uA 50Hz 9.3K 459.8
13788 LD2 10uA 40Hz 8.1K 471.6