• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name Ramesh/Diancheng

    Log entry time 15:54:02 on November 07, 2009

    Entry number 298114

    keyword=Shower missing channels

    Shower (it is in RHRS) PMT #8 signal is now plugged in and has signal;
    the input where the signal from PMT #8 is supposed to go was terminated
    by 50 ohm earlier.

    Shower PMT # 35 (histogram #34) had loose HV connection, fixed it now and
    it has signal.

    All shower signals now should appear in histograms.