User name Ramesh/Diancheng
Log entry time 17:05:40 on November 10, 2009
Entry number 298771
This entry is a followup to: 298638
keyword=bcm Dx3 check
Diancheng and I checked all BCM signals in counting house middle room by
looking on the scaler display. We unplugged one of the bcm signals and
observed that it disappears on the scaler display and vice-versa. All
signals respond correctly except d3. The signal d3 is on the scope but
never appears on the scaler display. We checked the scaler map, found
nothing wrong. There are two possible solutions:
1. The connection between counting house patch pannel to the detector
stack scaler is broken somewhere; need to check in next access.
2. May be we overlooked the real channel for d3 in LeCroy Module in
detector hut; need to recheck in next access.