Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name flay
Log entry time 07:55:46 on November 11, 2009
Entry number 298958
keyword=11/11/09 Wednesday Owl Shift Summary
00:00 Start of shift running smoothly taking production data
at I ~ 100 uA
04:00 Production data taking going smoothly.
05:45 Beam trip causes target temperature and pressure to drop.
Same problem as last owl shift (11/10/09). Coda runs stopped.
See HALOG 298938
07:00 Production data taking running smoothly.
07:27 Beam trip again causes target temp. & P to drop.
We've decided this time to let the coda runs continue,
as the target recovered quickly.
07:50 Last run set started: 13881/25385/4938. Production running
smoothly.
List of Runs (PVDIS/LHRS/RHRS):
Production:
13872/25375/4927 -- Stopped early because PVDIS DAQ froze; rebooted.
Junk runs.
13873/25376/4928
13874/25377/4929 -- Stopped early. Kai changed the trigger prescales.
See HALOG 298899.
13875/25378/4930
/25379/4931
/4932
13876/25380/4933
13877/25381/4934
13878/25382/4935 -- Stopped early. Beam trip caused the target pressure
to drop (caused by a temperature drop) -- same issue
as seen during last owl shift (11/10/09).
See HALOG 298938.
13879/25383/4936
13880/25384/4937 -- Beam trip causes target temp & P to drop.
We decided to leave the Coda running, as the
trip was short and the target recovered quickly.
13881/25385/4938