    User name jpchen

    Log entry time 13:55:55 on November 12, 2009

    Entry number 299222

    This entry is a followup to: 299082

    keyword=

    I was puzzled initially from the power chart that HP heater stayed at ~200 watts, which caused the target temperature to go up above 22K. Then I realized that although I turned the local heater power to zero (Halog 298543), somebody turned it back). I asked John to check and it turned out that the local power was at around 200 watts. So I asked John to turn it to zero. Target operator, please turn the local heater to 0 after each use (for ioc rebooting or for stabilizing temperature).