keyword=Compton table scan
Looking at the analyzed data:
The vertical table position is 38.5mm and the horizontal table position is 72mm +/- 1mm for the GSO to be centered on the photon beam. The nominal positions are y=43.75mm and x=73.5mm, so we need to move down in y by 5mm (consistent with old results), and move down in x by 1mm (during HAPPEX we centered the collimator in x to 73.5, so this is a slight shift from before). If we use a smaller collimator, we will need to move the collimator to be centered at this new position.