• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name R.M. et.al.

    Log entry time 14:39:07 on November29,2009

    Entry number 303537

    Followups:

    keyword=Compton tests, part I

    Will attempt to do low-I Compton simultaneously with Moller.

    Persons doing the test:

    A. Camsonne, M. Friend. D. Wang, R. Michaels, O. Glamazdin,
    M. Dalton

    13:50 Edet at 40 mm with 100 uA. Current drawn acceptable,
    approx 300 to 400 nA.

    14:00 The beam current was reduced to ~1 uA and Moller tune
    established.

    14:10 Alex instructed MCC to adjust IPM1P02B.YPOS to achieve
    maximum interception with laser.

    14:20 Beam current 0.8 uA. Tuning on Moller target.

    14:37 After some struggle, it's realized that making a tune
    for both Moller and Compton is impossible. Abandon Moller.
    Will continue with Compton test.