    keyword=LHRS dipole reset smoothly and is ramping back up

    Just for reference. I turned the regulation on D1 off, waited for the dipole to ramp down to 0A (takes a while), clicked on '0 Amps', waited, clicked on reset, waited for the fault to clear (it did), entered 0 into the current set field, clicked 'On' and watched the 'Power supply' and 'Magnet' indicators cycle through to ON and ENGAGED. I then ramped the magnet to 100A as a check, and turned the regulation back on to let the system take it back to the P0 setpoint.