User name R. Michaels
Log entry time 08:42:35 on December 19, 2009
Entry number 308134
This entry is a followup to: 308083
keyword=re: left arm scalers
If you look at any end-run entry in halog for Left HRS runs you'll
see that the scaler is reading -1 on all channels. This symptom of
the SIS3800 scaler means you have to power-cycle the scaler. Some-
times it's enough to reboot the crate but apparently that was tried.
And I don't think you can say "xscaler is not working". xscaler is
a GUI that displays the values. It is probably still working (I'm
not able to check it from here, though.) It's the scaler itself
which has a problem. The code that fills the halog is an independent
code. Another independent code is "scread". You can do the following
on a-onl to see -1 everywhere : "cd ~a-onl/scaler/left ; scread".
It's probably not important enough to interrupt the experiment -- we're
not measuring cross sections and we have rate info from the PVDIS DAQ.
A copy of this log entry has been emailed to: rsubedi,pkpan