NEXT
Make New Entry,
Make Followup Entry
User name A. Camsonne
Log entry time 17:51:17 on March 22, 2010
Entry number 313046
keyword=N1470 HV control instructions
To control the GEM HV,
one has to log into the ahut computer using the adaq account and use minicom.
Once in Minicom,
type CAEN and carrier return.
The menu will display, you should get :
## ## ## #### ###### #####
### ## #### ## ## ## ## ##
## # ## ## ####### ## ## ##
## ### ## ## ## ## ##
## ## #### #### ## #####
C.A.E.N. N1470 4 Ch HV Power Supply V1.00 Addr 00
B O A R D M E N U
Display Display/Modify channels
Format Reformat EEPROM
General General board status
Update Firmware Update
Quit
Select Item
---------------
Then type D to go in the control screen :
C.A.E.N. N1470 4 Ch HV Power Supply V1.00 Addr 00
Ch0 Ch1 Ch2 Ch3
Polarity - - - -
Vmon 0000.0 V 0000.0 V 0000.0 V 0000.0 V
Imon 000.000 uA 000.000 uA 000.000 uA 000.000 uA
Status PwF PwF PwF PwF
Power Off Off Off Off
Vset 3750.0 V 4200.0 V 4200.0 V 0200.0 V
Iset 0310.00 uA 0100.00 uA 0310.00 uA 0310.00 uA
Maxv 3800 V 8100 V 8100 V 8100 V
Ramp Up 050 V/S 050 V/S 050 V/S 050 V/S
Ramp Down 050 V/S 050 V/S 050 V/S 050 V/S
Trip 010.0 S 010.0 S 010.0 S 010.0 S
Power Down Kill Kill Kill Kill
Imon Range Low Low Low Low
Group Mode Reset Alarm Quit
and use the arrows to modify values. And space to toggle things.
A copy of this log entry has been emailed to: riordan@jlab.org,doug@jlab.org,nilanga@jlab.org,cisbani@jlab.org,urcioli@jlab.org,ole@jlab.org,camsonne@jlab.org