User name R. Michaels
Log entry time 00:13:01 on April26,2010
Entry number 319517
Followups:
keyword=Switch to thick C12
Will run with thick C12 tonight with rampdelay (effective) = 400 usec.
The actual parameters are
setTimeHAPTB(112, 1420);
Hence 112*2.5 usec + 120 usec = 400 usec
where 120 usec comes from 100 usec Tsettle plus
observed 20 usec additional delay.
The timing was checked on a scope in CH. Ramp delay 412 usec
and the Tend was about 200 usec. Fine.
Adjusted HV for thick C12. Need to check with panguin.
I'm a little worried about the swing shift data because
the detector HV seemed low to me, and the shift reported
that all prompt analyses are crashing now --> no plots.
R-HRS card 14
chan 4 = 950
chan 6 = 1000
L-HRS card 2
chan 2 = 865
chan 4 = 975
Also adjusted the Blumi in accord with the opmanual.
(The main detector HV for thick C12 are wrong in the opmanual.)
Run 3141 is junk. We are adjusting HV and checking signals
during this.