We moved the NMR from the Left Arm Dipole to the Right Arm Dipole. That seems to have stabilized the readback of the Right Dipole field. We installed the spare NMR on the Left Dipole. For the Right Q1, we tried installing the spare DAC (which was found in one of the Dipole spares drawers), but this did not solve the "bad bit" problem we were seeing. Setting the Q1 to 32A we got 5A out. We put the original DAC back in. We replaced the "regulator" card with our spare. We do not see the "bad bit" problem. So, the RQ1 seems to be ok for the moment. I found the LQ3 showing 1.25V output at 433.6A and the current control and voltage control LEDs switching like mad. I ramped down the magnet and locked the supply out. For the Left Dipole, we now have an additional problem with a fault preventing us from turning the supply ON. We quit for the day and will return to it tomorrow. So, the Right Arm seems functional and the Left Arm is down.