NEXT
Make New Entry,
Make Followup Entry
User name A. Camsonne
Log entry time 21:05:43 on June 11, 2010
Entry number 325773
keyword=Q2 measurements to be done again
I looked closer at the left arm runs, since they had trouble to generate
plots and saw that the Left Arm Quarts looked like a pedestal.
The plot for the right arm were generated ok but with low statistics.
It just came to my mind that the HV had to be increased for the PREX
detectors for the counting DAQ so we will retake the Q2 runs on Carbon
and thin lead.
I guess we can bill the 3 hours of beam to the GEM. I think the HV are
set up well for them. I will do a peaking time scan to check if the
timing is optimal.