    User name LeRose

    Log entry time 18:30:23 on July 1,2010

    Entry number 329315

    Followups:

    keyword=Right Arm Set

    The right HRS is now set for 1.131 GeV/c
    There were problems getting the NMR to lock for the right dipole. The solution was to, after typing in the desired P0, change the NMR settings on the NMR control screen as follows.
    Select probe A
    set the compensating coil voltage to 4.0 Volts (the original setting is too low).
    See the pictures attached


    FIGURE 1