User name R. Michaels
Log entry time 09:49:16 on July 02, 2010
Entry number 329406
This entry is a followup to: 329358
keyword=ADCs all working now
The front-panel gate on ROC1 slot 24 (and all others) is working now.
There was a bug in the programming of the module: two lines transposed.
Before programming one must address the module; these were flipped.
Oops !
A copy of this log entry has been emailed to: camsonne@jlab.org