    User name S.Abrahamyan

    Log entry time 06:01:18 on July 07, 2010

    Entry number 330604

    Followups:

    keyword=Shower/PreShower calibration

    After adjusting HV on Shower/PreShower using cosmics today the
    calibration of total absober in good shape.
    I analyzed run 1823 with all coefficients in DB set to 1.00.
    Results for that are shown on Figures 1,2 and 3
    Fig 1 --- Sh_E vs Ps_E 
    Fig 2 --- Sh_E vs Ps_E with cut on Gas Cherenkov(select elecrons)
    Fig 3 --- (1.94*Sh_E+Ps_E)/P with cut on Gas Cherenkov(select electrons)
    
    After that I used minimization algorithm to find coefficient. Run 1823
    was analyzed again with new coefficients from minimization. Results are
    shown on Figures 4,5 and 6
    Fig 4 --- Sh_E vs Ps_E 
    Fig 5 --- Sh_E vs Ps_E with cut on Gas Cherenkov(select elecrons)
    Fig 6 --- (1.94*Sh_E+Ps_E)/P with cut on Gas Cherenkov(select electrons)
    
    
    Resolution with pesent HV settings with coefficients = 1.0 is 9.3%
    After applying coefficients 6.8%.
    
    


    A copy of this log entry has been emailed to: bogdanw@jlab.org