• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name J. Zhang

    Log entry time 15:06:53 on July10,2010

    Entry number 331330

    This entry is a followup to: 331318

    keyword=Right Q1 was not recover to the previous after fixed

    After replaced the power supply board in the morning, the right Q1 field was not recovered to the value (0.01414) before 06:30 July 10, 2010. The filed value was about 0.01402. When it tripped at 14:18, it was set to 0.01398 until we recognized the difference. Then we set it back to 0.01413 at 14:30. Runs between 2010 and 2020 may be affected by this issue.


    FIGURE 1

    FIGURE 2