    keyword=Current scan for lumi detector

    The plots shows the result of the current scan done yesterday. 
    
    First canvas shows raw lumi rates vs beam current, they scale very well. 
    
    Second canvas shows the lumi rate(normalized by the BCM u10 rate) versus
    beam current. Below ~3uA the it is not linear. I will change HVs and do
    this test again to see if this improves at low current. Also, BCMs
    themselves may not be linear at below 2uA.
    
    