keyword=Cerenkov PMT1 now showing a nice pe peak
After raising the HV for cerenkov pmt 6 by a total of 50 V, that pmt now shows a nice pe peak.
I updated the db to reflect the new pe peak location.
Below are the calibration constants for the cerenkov. Maybe we will need to adjust the HV again to match the pe before using the cerenkov in the trigger.
calibration constants:
0.992 1.300 0.936 0.968 1.246 1.288 1.304 1.108 0.944 1.018
HV:
1752, 1800, 2000, 1571, 1926, 1651, 1650, 2000, 1840, 1749
The plot is from run 7124 (Sunday Oct 10, 3pm)
FIGURE 1