    User name jroche

    Log entry time 23:17:06 on October10,2010

    Entry number 334323

    This entry is a followup to: 334120

    Followups:

    keyword=Cerenkov PMT1 now showing a nice pe peak

    After raising the HV for cerenkov pmt 6 by a total of 50 V, that pmt now shows a nice pe peak.
    I updated the db to reflect the new pe peak location.

    Below are the calibration constants for the cerenkov. Maybe we will need to adjust the HV again to match the pe before using the cerenkov in the trigger.
    calibration constants:
    0.992 1.300 0.936 0.968 1.246 1.288 1.304 1.108 0.944 1.018
    HV:
    1752, 1800, 2000, 1571, 1926, 1651, 1650, 2000, 1840, 1749


    The plot is from run 7124 (Sunday Oct 10, 3pm)


    FIGURE 1