• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name Hisham

    Log entry time 16:08:43 on October18,2010

    Entry number 335490

    This entry is a followup to: 335029

    keyword=Correction: BCM calibration

    the previous entry (335482), did not take the double polsing clk into account. the new numbrs are:

    Tentative initial BCM calibration analysis data
    MCC log file 160,
    9 current steps 94, 50, 20, 10, 5, 2, 1, 0.5, 0.2 micro A
    Coda Data files 7353, 7354.

    These number are the slope of cnts vs clkcount
    clkcount is the 1024 kHz clock that is double pulsing at 2044 Hz.

    Columns are u1, u3, u10, respectively

    57.217 179.378 234.879
    30.483 93.900 142.303
    12.782 39.038 59.278
    6.402 19.375 29.374
    3.243 9.364 14.545
    1.369 3.874 5.772
    0.785 2.076 3.031
    0.476 1.123 1.579
    0.269 0.572 0.729

    d1, d3, d10

    62.910 193.395 225.142
    33.517 103.696 161.170
    13.943 43.228 136.156
    6.962 21.440 67.465
    3.475 10.634 33.398
    1.412 4.231 13.220
    0.763 2.237 6.882
    0.426 1.179 3.584
    0.229 0.565 1.647

    OLO2 current from MCC file 160 (halog 335002)
    http://www.jlab.org/~adaq/halog/html/1010_archive/101015110155.html
    Upstream, Downstream, repectively

    93.51 93.09
    49.03 49.50
    20.64 20.65
    10.21 10.21
    5.09 5.09
    20.022 2.021
    1.056 1.056
    0.54 0.54
    0.248 0.249

    Fit range: 10-100, 3-30, 1-15, respectively.
    Hisham




    FIGURE 1

    FIGURE 2