    User name efuchey

    Log entry time 23:23:44 on October18,2010

    Entry number 335536

    Followups:

    keyword=Results of QWP scan from 10/15/2010

    Right state:

    ####### RESULTS ########
    delta=-0.84328 +/- 0.0255645 degree
    Ellipse parameters:
    Axis Ratio=-0.665845 +/- 0.000747277
    ==> DOCP from ellipse param=-92.2638 +/- 0.0399348 %
    ==> DOCP from Stokes param=-92.0863 +/- -0.0740215 %
    Angle=79.1173 +/- -0.104971 degrees
    ==> retardation error=3.55372 +/- 0.151001 %

    left state:

    ####### RESULTS ########
    delta=20 +/- 0.000175184 degree
    Ellipse parameters:
    Axis Ratio=0.26361 +/- 0.00053417
    ==> DOCP from ellipse param=49.2964 +/- 0.0869114 %
    ==> DOCP from Stokes param=40.5875 +/- 0.0724453 %
    Angle=59.5394 +/- -0.0587952 degrees
    ==> retardation error=-6.3662 +/- 0.000286746 %

    As you see, there are severe problems with the left state, which comes from a misfollowing of the procedure