    User name miham

    Log entry time 22:32:14 on November 16, 2010

    Entry number 340408

    keyword=Q3 magnet trip

    Q3 magnet tripped. The current in the magnet went to zero. 
    In the Magnet Controls window we noticed a interlock fault.  
    We believe that the interlock was caused by the  fluctuation 
    in the He level. The level dropped to the red region for a 
    second and interlocked the magnet. We were able to reset 
    the controls. First we ramped the current up to 1600A. We stayed 
    there for 2min. Then we set it back to its production setting 
    of 741.0A. Because of this problem we had to stop run #8811.
    