    This entry is a followup to: 341727

    keyword=BCM u/d readings

    After calibration by Julie, the downstream BCM (yellow) is systematically reading roughly 1% higher current than the downstream (red). It might be a small drift since the calibration last Wednesday, or some inaccuracy in the calibration coefficient.

    The charge reported in the end-of-run file also shows this 1% difference, but I'm not sure if this has been calibrated yet.


    FIGURE 1