• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name saw

    Log entry time 16:21:10 on February 04, 2011

    Entry number 346065

    keyword=Terminators removed from scales in TS crate

    The 100 ohm termination resistor packs have been removed from all input channels in scales 2,3 and 4 in the trigger supervisor crate bbts1. The resistor packs are sitting on the ledge in side the TS crate in from of the VME modules. This is being done so that the DE and E signals can be daisy chained past the scalers to the F1 modules. This will give us independent TDCs on the DE and E tubes in case we have trouble with the 1190 TDC.