    keyword=ADC vs TDC in BigBite for T3 and T6 After Timing Setup

    Good news and bad news.  The timing adjustment worked well enough that we 
    are definitely catching the elastic protons with T3.  However, the retiming
    is not working quite right.  There is a 50 ns difference between the TDC
    for T3 and T6, shown in red and blue, respectively.
    
    For the elastic protons we only really care about T6, so we can probably
    live with this for tonight, but we will need to revisit the BigBite 
    retiming to fix this issue.
    


    A copy of this log entry has been emailed to: brads,gron,nmuangma,doug,kelleher