NEXT
Make New Entry,
Make Followup Entry
User name danez
Log entry time 15:20:30 on February 15, 2011
Entry number 347042
This entry is a followup to: 347024
Followups:
keyword=RHRS Shower Update - Disconnected Signal Cables
While investigating some of the missing signals in the right arm's shower
and preshower detectors, it was noticed that many of the cables connected
to the patch panel near the hut doors, which are then directly connected
to the ADCs, are not connected to anything in the patch panel near the
rear of the hut. That is, the patch panels do not match, so there are no
signals going into many of the TA's ADCs.
On Thursday, during beam studies, the signal cables will be tested with
an oscilloscope and reconnected properly. Additionally, depending on
time, the preshower cables in the right and the pion rejector cables in
the left arm will be checked.
A copy of this log entry has been emailed to: yez