User name A. Camsonne
Log entry time 09:29:33 on February 17, 2011
Entry number 347326
This entry is a followup to: 346957
keyword=VDC U1 check RHRS Run 1481
I swapped the connectors from Slot 13 channel 48 to 64 with slot 14
channel 48 to 64 to check if the cables or TDC are bad for for the 64
missing channels of U1.
TDC of slot 14 seems to gate gate fine and analyzer crate map looks fine.
This affects run 1481