User name V. Sulkosky
Log entry time 22:05:17 on February23,2011
Entry number 348345
keyword=VDC vs. S1 and S2m LHRS
In Figure 1 I have plotted the hit bar in S1 versus wire in u1 and
Fig. 2 shows hit bar in S2m versus wire in u1. The acceptance hole
does not appear to be a trigger related problem. But I did notice
that bar # 5 of S2m has lower counts then it's neighbors. This would
seem to indicate an inefficient trigger for this bar. Maybe HV related or a real issue with the trigger for bar 5.
The plots are from run 1782.
FIGURE 1
FIGURE 2