    User name V. Sulkosky

    Log entry time 12:04:06 on February 24, 2011

    Entry number 348436

    keyword=Found Stop for ROC 10 was inverted

    I was checking the trigger inputs to the TDC in ROC 10 and I noticed that
    the signals are coming about 300 ns before the stop.  However the data
    indicates that they are coming closer to 600 ns before the stop.  Kalyan 
    helped me verify that the stop signal going into the back of ROC 10 was
    indeed inverted.  
    
    I flipped the cable to be correct so I now do see that the trigger signals
    in ROC 10 come about 300 ns before the stop.  This change should make the
    DBB.t3, DBB.t1 variables make more sense.
    
    I am not sure how this impacts the wire chamber data besides the T0's will
    shift.  Hence they will need to be recalibrated.
    
    