• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name R. Michaels

    Log entry time 11:48:48 on February28,2011

    Entry number 349182

    keyword=scaler checks, inclu deadtime

    This is a detailed check of scalers recorded in the datastream
    for run 2036. They look fine. See rates from triggers, BCM,
    and other things (fig 1). The deadtime is computed differentially
    and assumes that every T6 is also a T3. This gives 14% deadtime
    (bottom, right panel) with fluctuations of about 1 or 2%.
    The result agrees with the deadtime displayed in the end-of-run
    entry for run 2036.

    So far, all is well in scaler land.



    FIGURE 1