    keyword=BCM calibration against OLO2 and OLO2 against Faraday Cup

    Using the data from BcmLog_170 and BcmLog_173.

    The OLO2 is calibrated using the absolute current from Faraday Cup (FC).

    I(OLO2) = [1.00202+/- 0.000225]*I(FC) +[0.01854+/-0.007383]

    The BCMs are calibrated using the OLO2.

    I(BCM1) = [0.01214]*I(OLO2)+[0.000288]; I(BCM2) = [0.01017]*I(OLO2)+[2.18E-5];

    see attachments.