    keyword=BCM cavity constants

    Using run#2254

    The current and the BCM scaler has the following relation:

    I_a = [N_a/t - offset_a]/const_K_a

    where I_a is the current

    N_a/t = scaler rate (Hz)

    off_set_a = off_set

    a = 1,3,10 is the gain factor

    we have:

    bcm constK Offset constK_err offset_err

    bcm_u1 2059.225227 627.980503 0.317307 10.325868

    bcm_u3 6385.086733 259.501488 0.317307 10.325868

    bcm_u10 19352.327633 627.422835 1.358402 12.284888

    bcm_d1 1234.216156 154.756427 0.317307 10.325868

    bcm_d3 7635.017957 636.693049 0.317307 10.325868

    bcm_d10 24207.080456 686.709379 1.358402 12.284888

    upstream: k10/k1 = 9.397868

    upstream: k3/k1 = 3.100723

    downstream: k10/k1 = 19.613323

    downstream: k3/k1 = 6.186127

    The downstream seem to be double the rate.

    k1_u/k1_d = 1.668448

    k3_u/k3_d = 0.836290

    k10_u/k10_d = 0.799449



    A copy of this log entry has been emailed to: vasulk,doug