User name Fatiha
Log entry time 17:23:44 on March13,2011
Entry number 351613
Followups:
keyword=LHRS D1-NMR not locked while its current looks fine: what you should do?
If it happens that D1 of Left HRS goes to zero while its current
looks all fine: go to Fields and click on the D1 NMR.
Turn search OFF, then set a value of DAC to a value close to
2200 (as shown in this case. might be different for other cases)
Now, it shouldn't take more than a couple of minutes for NMR to lock.