Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name LeRose
Log entry time 17:22:30 on March20,2011
Entry number 352689
This entry is a followup to: 352685
Followups:
keyword=Left Dipole NMR Lock
I get a better signal with 8.5 Volts. Presently the meter is in "AUTO" mode, "NEG"polarity, search "OFF", and DAQ=2121 (see attached).
If I switch to mode = "MAN" and set DAQ=2108 I get a signal on the o-scope that's 3 times larger at 8.5 Volts than at 6.5 Volts and switching to "AUTO" gets a lock at 8.5 Volts but not at 6.5 Volts
Regulation is on and holding.
I recommend:
keep DAC=2108, Comp Coils=8.5 Volts, mode="AUTO", Field Pol="NEG", and search="OFF"
FIGURE 1