• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name LeRose

    Log entry time 17:22:30 on March20,2011

    Entry number 352689

    This entry is a followup to: 352685

    Followups:

    keyword=Left Dipole NMR Lock

    I get a better signal with 8.5 Volts. Presently the meter is in "AUTO" mode, "NEG"polarity, search "OFF", and DAQ=2121 (see attached).
    If I switch to mode = "MAN" and set DAQ=2108 I get a signal on the o-scope that's 3 times larger at 8.5 Volts than at 6.5 Volts and switching to "AUTO" gets a lock at 8.5 Volts but not at 6.5 Volts
    Regulation is on and holding.
    I recommend:
    keep DAC=2108, Comp Coils=8.5 Volts, mode="AUTO", Field Pol="NEG", and search="OFF"


    FIGURE 1