Main INDEX
Monthly INDEX
PREV
NEXT
Make New Entry,
Make Followup Entry
User name yez
Log entry time 14:33:34 on April21,2011
Entry number 356375
keyword=T3,T3,T7 connected to R-HRS Scaler
Vince brought copies of T3,T7 from Bigbite weldment to the Right arm scaler, and I send the T4 signals from LHRS thought #5 fat cable to the Right arm scaler (through an ECL fanout ).
Basically what I did is using a discriminator to split the T4 signal that goes into Left Scaler into two copies, one of which goes into the scaler, and the other of which connects to the connector of #5 fat cable on the Left arm. On the right arm, a cable connects the T4 signal from the #5 fat cable to the most left channel of the patch panel #c3, then the signal is sent into the right arm scaler (through a level translator and an ECL fanout).
Vince will write a halog to summarize what he did.
FIGURE 1
FIGURE 2