    User name johna

    Log entry time 18:42:02 on April24,2011

    Entry number 356850

    This entry is a followup to: 356846

    Followups:

    keyword=HRS-R HV problems (and left arm monitoring problem)

    The right arm HV looked OK and there were no alarms, but it appeared as though the HV interface wasn't updating. It appeared to respond (e.g. when HV was turned off and on), but the voltatges weren't updating. Restarted the interface and turned on HV and we seem to be OK now, and triggers started up again.

    All of this was from the Right Arm HV window. Along the way, we lost connection to the LeCroy HV, which no longer provides any information on the voltages. Triggers seem to be OK, so we're running with both arms again, but need to check on the data quality (and see if we can reconnect to the HV controller)