NEXT
Make New Entry,
Make Followup Entry
User name V. Sulkosky
Log entry time 12:26:42 on May 08, 2011
Entry number 358604
This entry is a followup to: 355790
keyword=Restored SRC Triggers at Trigger Supervisor and BB Scalers
I restored the SRC triggers at the trigger supevisor. The 3-of-3 triggers
were disconnected from the T1 and T3 inputs and the normal 2-of-2 triggers
were restored as T1 and T3. I then connected BB and HAND singles to T6 and
T7 and reconnected HRS-coincidence to T5:
T1: right HRS (S1&&S2)
T2: right HRS efficiency
T3: left HRS (S1&&S2)
T4: left HRS efficiency
T5: LHRS&&RHRS coincidence
T6: BB singles
T7: HAND singles
T8: 1024 Hz clock
The datamon and BB scaler displays are correct. At the next access after
the pass change, I need to fix the mapping of the scalers in the
left and right HRS.
A copy of this log entry has been emailed to: camsonne,zhanxh,eip@tauphy.tau.ac.il,sgilad@mit.edu,doug,or.chen@mail.huji.ac.il,igor.korover@gmail.com,nmuangma