    User name V. Sulkosky

    Log entry time 20:29:15 on March 2,2012

    Entry number 362761

    Followups:

    keyword=Left HRS VDC Gas Flow

    Earlier I noticed that the left HRS bottom VDC gas flow was zero and
    sometimes the top chamber gas flow was also low. I talked with Jack
    and he said there was a problem with the read backs. The shift
    workers should pay attention to the right HRS VDC gas flows and the
    Gas Shed pressures. Also if the gas flow is disrupted, the Detector
    Gas System Interlock Status box will go into alarm, which is
    impossible to ignore when in the Counting House.

    Also new blubbers were recently installed in the two HRS huts. These
    can be checked during accesses to verify that the gas flow is still
    flowing. Perhaps at some point, we can get cameras installed so that
    the shift workers and routinely monitor the bubblers.