    keyword=[Target] Anneal Done

    Anneal finished, moved to Top after hitting ESC. Did anneal from hall as I had to plug in the Bypass valve controller cable.

    The anneal went MUCH smoother than usual, with the Top at ~88K and the Bot ~86K for 25 minutes. I attribute the closer temperatures to flow through the run valve instead of the bypass. I set the run to 1.5 and bypass to 0.3. The flow also stayed lower, near 30. Going up, the top stayed a few degrees below the bottom, which stayed a few degrees below the lowest resistor. Once I reached 85K, the top overtook the bottom, but not by the extent we're used to. I'll post plots later.