• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name jmaxwell

    Log entry time 18:41:55 on May 13, 2012

    Entry number 384692

    keyword=[Target] Anneal Plot

    Attached is a plot of today's anneal. If you compare it with some of the other here:

    http://twist.phys.virginia.edu/~jdm2z/g2p/anneals/

    you'll see the temperatures are much tighter. I achieved this using the bypass only opened to 0.3, the run opened to 1.3, and the pumping valves closed (because I forgot to open them). The top still kept creeping up, but this is the closest we've gotten to consistent by far.

    There's a bit of a hiccup where the separator runs out of liquid just before 16:10. I fixed it by opening up the separator flow to 30 and the temperatures came back together. I would keep the separator wide open in the future.






    Figure 1