User name R. Michaels
Log entry time 12:05:32 on November28,2012
Entry number 385739
keyword=status of L-HRS DAQ
Run the DAQ on the "adaq" account on adaql2.
The "test1" config is working; it is the TS11 crate only.
But this shows that the set up of adaq account is ok.
The "LHRS" config downloads and prestarts, but there are
no events. It is as if the RS485 cable is bad or the jumpers
not set correctly, or something.
What I think is the setup is below. (But I could be wrong
somehow). The HW roc numbers come from the jumper on the
TS RS485 line.
ROC4 ... top crate ... HW roc 4 ... hallasfi4 ... sfi4.crl
Has 9 TDCs in slots 3 - 11.
ROC5 ... middle crate ... HW roc 6 ... hallasfi5 .... sfi5.crl
Has 9 TDCs in slots 16 - 24.
ROC3 ... bottom crate ... HW roc 5 ... hallasfi3 .... sfi3.crl
Has 8 ADCs in slots 16 - 23.
All the ADCs and TDCs were tested with the usual Fastbus test
codes. They all work, except that slot 19 in ROC5 (middle
crate) has a strange response; it reports one header word.
We might want to swap that TDC. But I don't think that's why
I get no events.
Will keep working ....