    keyword=add delay to gates, see data now

    I added 650 nsec delay to the gates for LHRS DAQ. See the
    circuit in fig 1. This may not be optimal yet, but we need
    something like this.

    I had to add a discriminator after the delay, else the NIM/ECL
    would not fire.

    I also put two copies of the T1 into one of the TDCs and called
    them (temporarily) a V1 channel. I see these hits (fig 2), so
    this shows that the trigger + DAQ + software is working.

    I noticed that from the Hall A Tools GUI the VDC HV is off and
    the thresholds are off. Are we still controlling them this way ?
    Turning those on would lead to better results on the VDCs ...