• Main INDEX
  • Monthly INDEX
  • PREV
  • NEXT
    Make New Entry, Make Followup Entry

    User name rob

    Log entry time 15:40:34 on January 23, 2013

    Entry number 385769

    keyword=Compton e-detector: response to Sr source from three thick planes

    I used a Sr-90 beta source and an analog readout card to check for response from the three thin Si planes A,B,D in the Compton e-detector. Bias voltage was 80V (which is 160 ADC ticks in the GUI).

    Leakage currents at the start were

    A  1036  398
    B  1116  002
    C   000  000
    D  1505  001
    

    I saw no response from my cable for "plane D left", but I did not try more than one strip there.






    Plane A right



    Plane A left

    plane A left

    Plane B left

    plane B low left

    Plane B right

    plane B low right

    Plane D right

    plane D low right