I found I made the mistake with the Fastbus slot number they start at 0 instead of 1, so I had all my modules shifted in the crl. The missing modules were timing out hence no readout. I added ROC2,ROC1 and ROC6 to the db_cratemap, they were not there before hence nothing in histograms either. I advise to use the one in the $DB_DIR VDC cratemap was updated by Vince and looks good. DAQ and software should be ok now for VDC checkout.
Figure 1
Figure 2
Figure 3