    User name A. Camsonne

    Log entry time 21:40:06 on February 19, 2013

    Entry number 385792

    Followups:

    keyword=Electron detector source test : Compton electron detector chamber to not be open

    I disable most of the hot strips and lowered threshold. Put the amplifier
    discriminator board on plane C. With HV, noise seemed to be around 8 mV.
    Thresholds seems higher not sure if there is an offset. Will run with those.
    
    Made a small script to record the threshold in
    /home/compton/camsonne/script on comptom computer getthreshold.sh and
    SetThres.sh
    
    COM_DE_THRSET1                 20 
    COM_DE_THRSET2                 19 
    COM_DE_THRSET3                 22 
    COM_DE_THRSET4                 19 
    COM_DE_THRSET5                 18 
    COM_DE_THRSET6                 22 
    COM_DE_THRSET7                 22 
    COM_DE_THRSET8                 19 
    
    the file loaded is LowThres.txt
    
    with detector out from source there is 200 Hz of noise
    with detector lowered in front of the source there is about 800 Hz with
    single plane trigger.
    
    Started run 25156, with 2 out of 4 planes trigger, trigger rate is about 1 Hz
    
    
    


    A copy of this log entry has been emailed to: rob@jlab.org