keyword=Tests of 4th and 5th HV crates
Tests of the 4th and 5th of the refurbished HV crates.
Reminder: use halog 385829 as a testing procedure (follow up link)
E.g. for the 5th crate
Step 1 passed, see fig 1
Step 2 passed, but I had two problems:
a) port 2001 couldn't be used. Was not released by some other
process. I tried killing suspected processes, no joy. Rebooted
hacweb5, this worked.
b) Looks like you have to "cd" to the directory first and then
./LecroyHV_Shim
3. Step 3 worked, see GUI and see HV is on -- checked with voltmeter
------------
Reminder of some pitfalls
1. Some slots could be bad (I've seen it before). I think that if
the card's light flashes when you turn on the power, it's a good sign,
but if it doesn't flash it's a bad sign. (Not sure how well
correlated this is to functioning of slot, but I think 100%)
2. Remember to enter "-100" (with minus sign) if a N card.
3. I didn't check every channel of every slot. So, YMMV.
--------------
Serial numbers of LeCroy HV 1458 crates, and their status.
Serial num ........ Status
B52358 .... tested in the Spring
B11953 .... tested in the Spring
not recorded .... it was tested and is located in TEDF high-bay
B51981 .... tested just now
B69007 .... tested just now
B67561 .... staged, ready to test (perhaps tomorrow)
B52329 .... the same